![Rohde & Schwarz 1323.0012.03 User Manual Download Page 818](http://html1.mh-extra.com/html/rohde-and-schwarz/1323-0012-03/1323-0012-03_user-manual_1477777818.webp)
Command Reference
R&S
®
ZNL/ZNLE
818
User Manual 1178.5966.02 ─ 07
<CalKitLabel>
String parameter containing the label of a calibration kit available
on the analyzer, usually the serial number.
*RST:
n/a (A *RST does not change the assignment
between connector types and calibration kits.)
Example:
MMEM:LOAD:CKIT 'C:\Users\Public
\Documents\Rohde-Schwarz
\Vna\Calibration\Kits\New Kit 1
(123456).calkit'
Load the previously created cal kit file
New Kit 1 (123456).calkit
from the default cal kit direc-
tory.
CORR:CKIT:N50:LSEL 'New Kit 1', '123456'
Assign the imported kit to the N 50 Ω connector type, assuming
that the cal kit name stored in
New Kit 1 (123456).calkit
reads
New Kit 1
and that its label reads
123456
.
Manual operation:
"Add / Copy / Delete / Standards..."
[SENSe:]CORRection:CKIT:<StandardType>:WLABel
<ConnType>, <CalKitName>,
<CalKitLabel>, <StandardLabel>, <MinFreq>, <MaxFreq>, <DelayParam>,
<Loss>, <Z0>, <C0>, <C1>, <C2>, <C3>, <L0>, <L1>, <L2>, <L3>, OPEN |
SHORt | MATCh | <Resistance>[, <Port1>[, <Port2>]]
[SENSe:]CORRection:CKIT:<StandardType>:WLABel?
<ConnType>[,
<CalKitName>[, <CalKitLabel>[, <Port1>[, <Port2>]]]]
Defines the parameters of a non-ideal 1 port or 2-port calibration standard
<StandardType>
, where a particular cal kit can be addressed by name
and label
.
Apart from the additional
<CalKitLabel>
parameter, the syntax and semantics of this
command is identical to
Standard types and their parameters
.
Example:
CORR:CKIT:FOP:WLAB 'N 50 Ohm','New Kit
1','123456', '',
0,4000000000,0,0,50,99,-2.3,0.22,0,0,0,0,0
Define the properties of the Open (f) standard for a N 50 Ω con-
nector type in the calibration kit "New Kit 1" labelled "123456".
CORR:CKIT:FOP:WLAB? 'N 50 Ohm', 'New Kit 1',
'123456'
Query the properties of the Open (f) standard for a N 50 Ω con-
nector type in the calibration kit.
Manual operation:
"Add / Copy... / Delete / View / Modify..."
[SENSe:]CORRection:CKIT:<OnePortStandardType>:WLABel:SDATa?
<ConnectorType>, <CalKitName>[, <CalKitLabel>[, <PhysPort>]]
Reads the S-parameter data for a particular
1-port
cal kit standard previously loaded
from Touchstone file using
or
VNA Command Reference
www.allice.de
Allice Messtechnik GmbH