Command Reference
R&S
®
ZNL/ZNLE
885
User Manual 1178.5966.02 ─ 07
Example:
*RST; :SWE:STEP?
Query the default step size. Currently for all analyzers the
default sweep span is the full frequency range and the default
number of points is 201. Hence the response is (max. frequency
- min. frequency) / 200.
SWE:STEP UP
Increase the step size.
FREQ:STOP?; :SWE:POIN?
Query the stop frequency of the sweep and the number of
points. Increasing the step size has changed both values.
Manual operation:
[SENSe<Ch>:]SWEep:TIME
<SweepDuration>
Sets the duration of the sweep ("Sweep Time"). Setting a sweep time disables the
automatic calculation of the (minimum) sweep time; see
Note:
The sweep time is ignored for the sweep types "Time" and "CW Mode"
(
Suffix:
<Ch>
.
Channel number
Parameters:
<SweepDuration>
Sweep time. The minimum sweep time depends on the other
channel settings, in particular on the number of points
(
(
[SENSe<Ch>:]BWIDth[:RESolution]
ment delay for each partial measurement (
). Changing the duration leaves the number of
points unchanged but directly affects the delay.
Range:
Minimum sweep time to 100000 s.
*RST:
Minimum sweep time, depending on the channel
settings.
Default unit: s
Example:
SWE:TIME 1
Set a total sweep time of 1 s.
SWE:DWEL?
Query the delay for each partial measurement.
SWE:TIME 2
Increase the total sweep time to 2 s.
SWE:DWEL?
Query the meas. delay for each partial measurement again. The
delay is increased by 1 s divided by the total number of partial
measurements per sweep.
Manual operation: