![Rohde & Schwarz 1323.0012.03 User Manual Download Page 405](http://html1.mh-extra.com/html/rohde-and-schwarz/1323-0012-03/1323-0012-03_user-manual_1477777405.webp)
VNA GUI Reference
R&S
®
ZNL/ZNLE
405
User Manual 1178.5966.02 ─ 07
●
"Import File..." opens a dialog to load a limit line from a trace file (see
In case the selected file contains more than one trace, another popup dialog lets
you select the adequate one:
Imported traces are polygonal curves with n points and n – 1, where n is the "Number
of Points" of the imported trace (see
Chapter 9.5.1, "Sweep Params Tab"
on page 302). The n – 1 segments are appended to the current segment table for fur-
ther editing. Existing limit line segments are not overwritten.
Both import dialogs contain the following file import settings:
●
"Offsets" contains two input fields to define constant offset values for all imported
segments. The "Response" offset shifts all segments in vertical direction, the "Stim-
ulus" offset shifts them in horizontal direction. The offsets are added to the start
and stop values of all segments.
●
"Type" defines whether the imported segments belong to the "Upper" or "Lower"
limit line. A third option is to import the segments but disable the limit check ("Off").
Remote command:
CALCulate<Chn>:LIMit:LOWer:FEED
CALCulate<Chn>:LIMit:UPPer:FEED
9.9.5
Ripple Test Tab
Defines ripple limits for the measurement results, visualizes them in the diagrams and
activates/deactivates the ripple limit check.
Display Lines Softtool
www.allice.de
Allice Messtechnik GmbH