VNA GUI Reference
R&S
®
ZNL/ZNLE
440
User Manual 1178.5966.02 ─ 07
Use
●
"Active Channel" if you only want to calibrate the active channel
●
"All Channels" if you want to calibrate all channels in the current channel setup
●
"Selected Channels" if you want to calibrate only certain channels
9.12.1.3
Calibration Unit Wizard
The "Calibration Unit" wizard guides you through the setup and execution of an auto-
matic calibration.
Access:
[Cal] > "Calibration" > "Start... (Cal Unit)"
The wizard proceeds in three steps.
1.
allows you to configure the calibration and to select a calibration unit
(along with its characterization).
2.
allows you to define the port assignments between the
R&S
ZNL/ZNLE and the calibration unit.
3.
guides you through the required measurements on the standards
provided by the calibration unit. Finally, you can either apply the resulting error
terms to the related channels, or discard them.
Background information
Chapter 8.5.5, "Automatic Calibration"