Command Reference
R&S
®
ZNL/ZNLE
862
User Manual 1178.5966.02 ─ 07
[SENSe<Ch>:]FREQuency:STARt
<StartFreq>
[SENSe<Ch>:]FREQuency:STOP
<StopFreq>
These commands defines the start and stop frequency for a frequency sweep. The val-
ues also define the display range in a Cartesian diagram. The default start and stop
frequencies equal to the minimum and maximum frequency of the analyzer.
Suffix:
<Ch>
.
Channel number
Parameters:
<StopFreq>
Start and stop frequency of the sweep
Increment: 0.1 kHz
Default unit: Hz
Example:
*RST; FREQ:STAR 100000
Activate a frequency sweep and set the start frequency to 100
kHz.
FREQ:STOP 10MHz
Set the stop frequency to 10 MHz.
Manual operation:
"Start Frequency / Stop Frequency / Center Frequency /
Note:
If the start frequency entered is greater than the current stop frequency, the stop
frequency is set to the start frequency plus the minimum frequency span
(
).
If the stop frequency entered is smaller than the current start frequency, the start fre-
quency is set to the stop frequency minus the minimum frequency span.
[SENSe:]HARMonic...
Implements functions related to harmonic grids for time domain transformation.
..............................................................................................862
......................................................................................863
[SENSe<Ch>:]HARMonic?
Queries whether the current frequency grid is harmonic.
Suffix:
<Ch>
.
Channel number
Return values:
<arg0>
Usage:
Query only
Manual operation:
"Is the Current Grid Harmonic?"
VNA Command Reference
www.allice.de
Allice Messtechnik GmbH