VNA GUI Reference
R&S
®
ZNL/ZNLE
303
User Manual 1178.5966.02 ─ 07
System error correction
In general, the system error correction is no longer valid after a change of the sweep
parameters. The status of the calibration is shown in the trace list. If the number of
points is changed, the analyzer interpolates the correction data. The calibration label
"Cal Int" is displayed. See also
Chapter 8.5.4, "Calibration Labels"
Number of Points
Sets the total number of measurement points per sweep. The minimum number of
points is 1 (measurement at a single frequency/power/time value). The maximum
depends on the analyzer type.
Sets the total number of measurement points per sweep. The minimum number of
points is 1 (measurement at a single frequency/power/time value), the maximum is
100,001.
Together with the sweep range defined in the
of the "Stimulus" softtool,
this parameter defines the grid of sweep points. The step width between two consecu-
tive sweep points is constant on a linear scale (sweep types "Lin Freq", "Time" and
"CW Mode") or on a logarithmic scale (sweep types "Log Freq" and "Power").
The sweep points for linear frequency sweeps can also be defined using the
Tip: Measurement time and screen resolution
Increasing the number of sweep points improves the resolution of the trace but increa-
ses the measurement time. The overall measurement time is composed of the hard-
ware settling time at the beginning of the sweep plus the sum of the measurement
times at each individual sweep point. Hence the measurement time increases roughly
linearly with the number of points.
Chapter 8.1.4, "Sweep Control"
Remote command:
Freq Step Size
Sets the distance between two consecutive sweep points of a linear frequency sweep.
Sweep Softtool