VNA Concepts and Features
R&S
®
ZNL/ZNLE
252
User Manual 1178.5966.02 ─ 07
8.5.2.4
Cal Kit Files
Calibration kit files can be used to store the parameters of a particular calibration kit, to
reload the data and to exchange calibration kits from one network analyzer to another.
Cal kit file contents
Cal kit files are independent of the current channel setup and contain the following
information:
●
Name and label of the calibration kit
●
Connector type including all connector type parameters (name, polarity, offset
model, reference impedance)
●
Type, gender and label of all standards in the kit together with the circuit model
parameters (offsets, load) or S-parameter tables (.s<n>p file) that are necessary to
determine its magnitude and phase response.
By default cal kit files are stored in the
C:\Users\Public\Documents\Rohde-Schwarz\ZNL\Calibration
directory.
●
To
export
cal kit data, the analyzer uses a specific binary file format
*.calkit
.
●
Three different
import
file formats are supported: R&S
ZVA-specific binary cal kit
files (
*.calkit
), R&S
ZVR-specific binary cal kit files (
*.ck
), cal kit files in Agi-
lent-specific ASCII formats (
*.csv
,
*.prn
).
Importing older R&S
ZVR cal kit files
On loading some older R&S
ZVR-specific
*.ck
files, e.g. the R&S
ZV-Z23 cal kit file,
the R&S
ZNL/ZNLE generates the message "File does not comply with instrument cali-
bration kit file format". The files must be converted using an R&S
ZVR network ana-
lyzer equipped with a firmware version V3.52 or later.
Proceed as follows:
●
On the R&S
ZVR, press "CAL > CAL KITS > MODIFY KITS > INSTALL NEW KIT"
to import the
*.ck
file.
●
Press "CREATE INST FILE" in the same submenu to export the
*.ck
file in a
R&S
ZNL/ZNLE-compatible format.
●
Import the converted file into the R&S
ZNL/ZNLE.
*.csv cal kit files: VNA Cal Kit Manager V2.1
The VNA Cal Kit Manager is a free, Windows
®
-based software tool intended to import,
edit, and export *csv cal kit files. The software is available for download at
. The decimal separator used by the VNA Cal Kit Man-
ager V2.1 depends on the language version of the Windows
®
operating system: Cal kit
files generated on an English operating system contain dots, the ones generated on a
German system contain commas.
The network analyzer expects the dot as a separator and displays an error message
when a
*.csv
file with commas is loaded. Please install the VNA Cal Kit Manager
V2.1 on an appropriate (e.g. English) Windows
®
version to avoid trouble.
Calibration
www.allice.de
Allice Messtechnik GmbH