Command Reference
R&S
®
ZNL/ZNLE
596
User Manual 1178.5966.02 ─ 07
Manual operation:
See
CALCulate<Chn>:FILTer[:GATE]:TIME:SHAPe
<TimeGate>
Selects the time gate to be applied to the time domain transform.
Tip:
Use the generalized command
CALCulate<Chn>:FILTer[:GATE]:TIME:WINDow
if you wish to select a Dolph-Chebychev time gate.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<TimeGate>
MAXimum | WIDE | NORMal | MINimum
MINimum - Steepest edges (rectangle)
WIDE - Normal gate (Hann)
NORM - Steep edges (Hamming)
Maximum - Maximum flatness (Bohman)
*RST:
WIDE
Example:
*RST; :CALC:FILT:TIME:SHAP?
Reset the instrument and query the type of time gate used. The
response is WIDE.
Manual operation:
See
CALCulate<Chn>:FILTer[:GATE]:TIME:SHOW
<Boolean>
Enables or disables permanent display of the gate limits.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean>
ON - time gate permanently displayed
OFF - time gate hidden
*RST:
OFF
Example:
See
CALCulate<Chn>:FILTer[:GATE]:TIME:CENTer
Manual operation:
See
CALCulate<Chn>:FILTer[:GATE]:TIME:SPAN
<Span>
Defines the span of the time gate.
Suffix:
<Chn>
.
Channel number used to identify the active trace
VNA Command Reference
www.allice.de
Allice Messtechnik GmbH