SYSTem:COMMunicate:NET:HOSTname
<HostName>
Sets or gets the host name of the instrument.
Parameters:
<HostName>
Host name
SYSTem:COMMunicate:RDEVice:AKAL:ADDRess
<Address>
Selects one of the USB-connected calibration units for calibration (see commands
SENSe<Ch>:CORRection:COLLect:AUTO...
). This command is not necessary if
only one cal unit is connected.
Parameters:
<Address>
Name (USB address) of a connected calibration unit (string vari-
able). The names of all connected cal units can be queried using
SYSTem:COMMunicate:RDEVice:AKAL:ADDRess:ALL?
.
Example:
SYST:COMM:RDEV:AKAL:ADDR:ALL?
Query the names of all connected calibration units.
SYST:COMM:RDEV:AKAL:ADDR 'ZV-Z52::1234'
Select the cal unit named 'ZV-Z52::1234' for calibration.
CORR:COLL:AUTO '', 1, 2, 4
Perform an automatic 3-port TOSM calibration at test ports 1, 2,
and 4 using the calibration unit's default calibration kit file and
arbitrary test ports of the cal unit.
Manual operation:
SYSTem:COMMunicate:RDEVice:AKAL:ADDRess:ALL?
Queries the names (USB addresses) of all connected calibration units.
Example:
SYSTem:COMMunicate:RDEVice:AKAL:ADDRess
Usage:
Query only
Manual operation:
SYSTem:COMMunicate:RDEVice:AKAL:CKIT:CATalog?
Queries all characterizations (cal kit files) which are stored on the connected calibra-
tion unit, either on the calibration unit's internal (flash) memory or on an SD card inser-
ted at the calibration unit (prefix: "SD"). A possible response is
'Factory,ZN-Z51_custom,Throughs,SD:test'
. The factory characterization is
always available; an empty string denotes that no calibration unit is connected.
If several cal units are USB-connected to the analyzer, the command queries the cal
unit selected via
SYSTem:COMMunicate:RDEVice:AKAL:ADDRess
.
VNA Command Reference