Common Instrument Functions
R&S
®
ZNL/ZNLE
174
User Manual 1178.5966.02 ─ 07
To change the instrument's computer name
1. Press the [Setup] key and then the "N Remote" softkey.
The current "Computer Name" is displayed in the "Network" tab.
2. Enter the new computer name and close the dialog box.
The configuration is saved, and you are prompted to restart the instrument.
3. Confirm the displayed message ("Yes" button) to restart the instrument.
How to Configure the LAN Using the LXI Web Browser Interface
The instrument's LXI browser interface works correctly with all W3C compliant brows-
ers.
► In the web browser, open the
http://<instrument-hostname>
or
http://
<instrument-ip-address>
page, e.g.
http://10.113.10.203
.
The default password to change LAN configurations is
LxiWebIfc
.
The "Instrument Home Page" (welcome page) opens.
The instrument home page displays the device information required by the LXI stan-
dard including the VISA resource string in read-only format.
► Press the "Device Indicator" button on the "Instrument Home Page" to activate or
deactivate the LXI status LED to the right of the button. A green LXI status LED
indicates that a LAN connection has been established; a red status LED indicates
an error, for example, that no LAN cable is connected. When a device is connect-
Network and Remote Settings