Operating Modes, Applications, Channel Setups, and Result Displays
R&S
®
ZNL/ZNLE
93
User Manual 1178.5966.02 ─ 07
Figure 6-1: Defining Channel Setups (R&S
ZNL with R&S
ZNL-B1)
Switching between channel setups
When you switch to a new channel setup, a set of parameters is passed on from the
current channel setup to the new one:
●
Center frequency and frequency offset
●
Reference level and reference level offset
●
Attenuation
After initial setup, the parameters for the channel setup are stored upon exiting and
restored upon re-entering the channel setup. Thus, you can switch between measure-
ments quickly and easily.
...............................................................................................93
....................................................................................... 93
.................................................................... 94
Duplicate Current Channel Setup
.................................................................. 94
................................................................................................94
Defining a channel setup
To start a new channel setup or replace an existing one, select the corresponding
application in the "Mode" dialog box.
Note:
The channel setups are labeled with the application name. If that name already
exists, a sequential number is added. You can change the name of the channel setup
by double-tapping the name in the channel setup bar and entering a new name.
For an overview of default names see
Remote command:
New Channel Setup ← Defining a channel setup
The application selected on this tab of the dialog box activates a new channel setup,
i.e. a new tab in the display.
Defining Channel Setups