Command Reference
R&S
®
ZNL/ZNLE
527
User Manual 1178.5966.02 ─ 07
Example:
MMEM:SEL:ALL
//Save all items (User Settings, All Traces, All Limit Lines) from
the R&S
ZNL/ZNLE.
MMEM:LOAD:STAT 1,'C:
\Users\Public\Documents\Rohde-Schwarz\ZNL\user\TEST01'
//Reloads all items
In the "Recall" dialog, select only "User Settings" and "All Limit
Lines".
MMEM:LOAD:STAT 1,'C:
\Users\Public\Documents\Rohde-Schwarz\ZNL\user\TEST01'
//Reloads user settings and all limit lines.
*RST
//Reset instrument.
MMEM:LOAD:STAT 1,'C:
\Users\Public\Documents\Rohde-Schwarz\ZNL\user\TEST01'
//Selected items are retained. Reloads user settings and all limit
lines.
Restart the instrument.
(Switch the [ON/OFF] key off and on).
MMEM:LOAD:STAT 1,'C:
\Users\Public\Documents\Rohde-Schwarz\ZNL\user\TEST01'
// Selected items are set to default. Reloads only the user set-
tings.
Manual operation:
MMEMory:LOAD:TYPE
<Mode>
This command defines whether the channel setups that will be loaded with the subse-
quent
MMEM:LOAD:STAT
command will replace the current channel setup or activate a
new channel setup.
Parameters:
<Mode>
NEW | REPLace
NEW
The loaded settings will be activated in a new channel setup.
REPLace
The loaded settings will replace the currently active channel
setup.
*RST:
NEW
Common Instrument Commands
www.allice.de
Allice Messtechnik GmbH