Command Reference
R&S
®
ZNL/ZNLE
854
User Manual 1178.5966.02 ─ 07
*RST:
0
Manual operation:
"Offset > Calculate after De-/Embed."
[SENSe<Ch>:]CORRection:EWAVe[:STATe]
<Boolean>
Activates or deactivates the wave correction logic for wave quantities and ratios.
Suffix:
<Ch>
.
Channel number
Parameters:
<Boolean>
*RST:
ON (1)
[SENSe<Ch>:]CORRection:LOSS<PhyPt>
<OffsetLoss>
Defines the offset loss at the reference frequency (
).
Suffix:
<Ch>
.
Channel number of the offset-corrected channel
<PhyPt>
Port number of the analyzer
Parameters:
<OffsetLoss>
Frequency-dependent part of the offset loss
Range:
-200 dB to +200 dB
Increment: 0.001 dB
*RST:
0 dB
Default unit: dB
Example:
[SENSe<Ch>:]CORRection:EDELay<PhyPt>:ELENgth
Manual operation:
"Loss at DC / Loss at Freq / Freq for Loss"
[SENSe<Ch>:]CORRection:LOSS<PhyPt>:AUTO
<Activate>
Defines the offset parameters for the active test port such that the residual delay of the
active trace (defined as the negative derivative of the phase response) is minimized
and the measured loss is reproduced as far as possible across the entire sweep range
("Auto Length and Loss").
Suffix:
<Ch>
.
Channel number of the offset-corrected channel
<PhyPt>
Port number of the analyzer. This numeric suffix is ignored; the
active port is determined by the active trace.
Setting parameters:
<Activate>
ONCE
Applies the "Auto Length and Loss" function.
VNA Command Reference
www.allice.de
Allice Messtechnik GmbH