Command Reference
R&S
®
ZNL/ZNLE
864
User Manual 1178.5966.02 ─ 07
Example:
*RST; :CALC:PAR:DEL 'TRC1'
Reset the analyzer and delete the (default) trace for channel no.
1.
SOUR:LPOR1 1,2
Combine the physical ports no. 1 and 2 to define the balanced
(logical) port no. 1.
LPOR1:ZDIF 27, 2
Specify a complex differential mode reference impedance of
27Ω + j * 2Ω for the defined port.
Manual operation:
"Single Ended Mode / Common Mode / Differential Mode"
[SENSe<Ch>:]LPORt<LogPt>:ZDEFault[:STATe]
<arg0>
Allows to toggle between default and renormalized reference impedance(s) for logical
port <LogPt>.
Suffix:
<Ch>
.
Channel number
<LogPt>
Logical port number
Parameters:
<arg0>
ON (1): Use default impedance(s)
OFF (0): Use the redefined impedances defined via
[SENSe<Ch>:]PORT<PhyPt>:ZREFerence
for single-ended
[SENSe<Ch>:]LPORt<LogPt>:ZCOMmon
and
[SENSe<Ch>:]LPORt<LogPt>:ZDIFferent
for balanced
logical ports
Manual operation:
[SENSe:]PORT...
The
[SENSe:]PORT...
commands define the reference impedances at the physical
ports.
[SENSe<Ch>:]PORT<PhyPt>:ZREFerence
<RealPart>[, <ImaginaryPart>]
Specifies the complex reference impedance for the physical port numbered
<PhyPt>
(impedance renormalization).
[SENSe<Ch>:]LPORt<LogPt>:ZDEFault[:STATe]
to toggle between config-
ured and default reference impedances.
Suffix:
<Ch>
.
Channel number
<PhyPt>
Physical port number
VNA Command Reference
www.allice.de
Allice Messtechnik GmbH