Command Reference
R&S
®
ZNL/ZNLE
832
User Manual 1178.5966.02 ─ 07
<CalUnitPort1>
Port numbers of the cal unit. For an n-port automatic calibration,
n arbitrary (not necessarily consecutive) port numbers must be
specified. It is possible to combine arbitrary (not necessarily
matching) pairs of analyzer and cal unit ports.
<TestPort2>
Example:
CORR:COLL:AUTO:PORT '', 1, 2
Perform an automatic 1-port calibration at analyzer port 1, using
the calibration unit's default calibration kit file and port 2 of the
cal unit.
Usage:
Setting only
Manual operation:
See
[SENSe<Ch>:]CORRection:COLLect:AUTO:PORTs:TYPE
<CalType>[,
<Characterization>, <TestPort1>, <CalUnitPort1>, <TestPort2>...]
Selects and initiates an automatic calibration at arbitrary test and calibration unit ports.
A progress monitor for the calibration sweeps is displayed.
Tip:
This command is necessary if the analyzer fails to detect the cal unit ports con-
nected to each of its ports (e.g. because of a high attenuation in the test setup). If auto-
detection works you can use the simpler command
Suffix:
<Ch>
.
Channel number of the calibrated channel
Setting parameters:
<CalType>
FNPort | FOPort | UTRans | REFL | RSHort
FNPort - full n-port (UOSM) calibration
FOPort - full one-port ("Refl OSM") calibration
REFL - Reflection normalization, Open
RSHort - Reflection normalization, Short
<Characterization>
Location of the characterization (cal kit file) to be used for the
automatic calibration (string parameter):
–
If an empty string (
' '
) is specified, the factory cal kit file
stored in the active calibration unit is used.
–
A VNA cal kit file name
*.calkit
without path refers to a
specific cal kit file stored in the internal memory of the active cal-
ibration unit.
–
A VNA cal kit file name without path but prefixed with "SD:"
refers to a specific cal kit file stored on the SD card inserted at
the active calibration unit.
–
A VNA cal kit file name
*.calkit
with path refers to a spe-
cific cal kit file stored in an arbitrary directory on the analyzer.
VNA Command Reference
www.allice.de
Allice Messtechnik GmbH