CALCulate<Chn>:RIPPle:FAIL?
Returns a 0 or 1 to indicate whether or not the global ripple limit check has failed.
Tip:
CALCulate<Chn>:RIPPle:SEGMent<Seg>:RESult?
to query the result
for a single ripple limit range.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Example:
*RST; CALC:RIPP:DATA 1, 1500000000, 2000000000,
3
Define and enable a ripple limit range in the stimulus range
between 1.5 GHz and 2 GHz, assigning a ripple limit of +3 dB.
CALC:RIPP:STAT ON; FAIL?
Switch the limit check on and query the result.
CALC:RIPP:STAT:AREA LEFT, TOP
For a subsequent check at the GUI or a hardcopy, move the info
field to the top left position.
Usage:
Query only
Manual operation:
CALCulate<Chn>:RIPPle:RDOMain:FORMat
<UnitRef>
Deletes the existing ripple limit ranges and (re-)defines the physical unit of the ripple
limit. The units of the stimulus values are defined via
Suffix:
<Chn>
.
Channel number used to identify the active trace
Setting parameters:
<UnitRef>
COMPlex | MAGNitude | PHASe | REAL | IMAGinary | SWR |
GDELay | L | C
Keyword for the physical unit of the response values; dimension-
less numerss, relative power, phase, time, inductance, capaci-
tance units.
*RST:
n/a
Default unit: 1 (U, for COMPlex, REAL, IMAGinary, and SWR);
dB (for MAGNitude), deg (for PHASe), s (for GDE-
Lay), H (Henry, for L), F (Farad, for C).
Example:
*RST; CALC:RIPP:DATA 1, 1500000000, 2000000000,
3
Define and enable a ripple limit range in the stimulus range
between 1.5 GHz and 2 GHz, assigning a ripple limit of +3 dB.
CALC:RIPP:RDOM:FORM COMP
Delete the ripple limit range, select complex units for the ripple
limit.
Usage:
Setting only
VNA Command Reference