Common Instrument Functions
R&S
®
ZNL/ZNLE
149
User Manual 1178.5966.02 ─ 07
Geometric Calculation of Bandfilter Center
Defines how bandfilter searches calculate the center frequency of the passband or
stopband (see
If "Geometric Calculation of Bandfilter Center" is checked, the
geometric mean
of the
lower band edge and upper band edge frequencies is used, otherwise their
arithmetic
mean
.
Remote command:
CALCulate:MARKer:FUNCtion:BWIDth:GMCenter
Touchstone Export Options
Configures whitespace insertion during Touchstone file export.
The default export format is explained in
Chapter 8.4.2.1, "Touchstone Files"
●
logical columns are vertically aligned using spaces
●
positive and negative numbers are vertically aligned by prefixing positive numbers
with blanks
●
the frequencies are horizontally separated from the corresponding S matrices
using leading spaces
●
the content parts (header, S matrices for different frequencies) are separated by
blank lines
Use TAB (instead of blanks) ← Touchstone Export Options
If checked, columns are separated by tabs rather than spaces.
Remote command:
MMEMory:STORe:TRACe:OPTion:TABS
Trim Leading Whitespace ← Touchstone Export Options
If checked, whitespace at the beginning of each line is removed.
Remote command:
MMEMory:STORe:TRACe:OPTion:TRIM
General Instrument Setup
www.allice.de
Allice Messtechnik GmbH