Command Reference
R&S
®
ZNL/ZNLE
594
User Manual 1178.5966.02 ─ 07
CALCulate<Chn>:FILTer[:GATE]:TIME:DCHebyshev
......................................................... 595
CALCulate<Chn>:FILTer[:GATE]:TIME:SHAPe
..................................................................596
CALCulate<Chn>:FILTer[:GATE]:TIME:SHOW
.................................................................. 596
CALCulate<Chn>:FILTer[:GATE]:TIME:SPAN
....................................................................596
CALCulate<Chn>:FILTer[:GATE]:TIME:STARt
...................................................................597
CALCulate<Chn>:FILTer[:GATE]:TIME:STOP
................................................................... 597
CALCulate<Chn>:FILTer[:GATE]:TIME:STATe
................................................................... 597
CALCulate<Chn>:FILTer[:GATE]:TIME:WINDow
............................................................... 598
CALCulate:FILTer[:GATE]:TIME:AOFFset
<Boolean>
Activates the operating mode where the time gate is moved in the opposite direction
when the "Delay" setting is changed.
Parameters:
<Boolean>
ON | OFF - enable or disable "Adjust Time Gate".
*RST:
OFF
Example:
*RST; :CALCulate1:TRANsform:TIME:STATe ON
CALCulate1:FILTer:GATE:TIME:STATe ON; SHOW ON
Activate time domain representation and a time gate in channel
no. 1. Display the time gate
CALCulate1:FILTer:GATE:TIME:STARt 2ns; STOP 3
ns
Restrict the time gate to the time interval between 2 ns and 3 ns.
CALCulate:FILTer:GATE:TIME:AOFFset ON
Activate an offset of the time gate according to a new delay set-
ting.
SENSe1:CORRection:EDELay1:TIME 1ns
Specify a 1 ns delay at port 1.
CALCulate1:FILTer:GATE:TIME:STARt?; STOP?
Query the time gate position. The response is
1E-009;2E-009
.
Manual operation:
CALCulate<Chn>:FILTer[:GATE]:TIME[:TYPE]
<TimeGateFilter>
Selects the time gate filter type, defining what occurs to the data in the specific time
region.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<TimeGateFilter>
BPASs | NOTCh
BPASs
Band pass filter: Pass all information in specified time region and
reject everything else
NOTCh
Notch filter: Reject all information in specified time region and
pass everything else
VNA Command Reference
www.allice.de