Command Reference
R&S
®
ZNL/ZNLE
515
User Manual 1178.5966.02 ─ 07
Manual operation:
INITiate:SEQuencer:IMMediate
This command starts a new sequence of measurements by the Sequencer.
Before this command can be executed, the Sequencer must be activated (see
Example:
SYST:SEQ ON
Activates the Sequencer.
INIT:SEQ:MODE SING
Sets single sequence mode so each active measurement will be
performed once.
INIT:SEQ:IMM
Starts the sequential measurements.
Manual operation:
INITiate:SEQuencer:MODE
<Mode>
This command selects the way the R&S
ZNL/ZNLE application performs measure-
ments sequentially.
Before this command can be executed, the Sequencer must be activated (see
Note:
In order to synchronize to the end of a sequential measurement using *OPC,
*OPC? or *WAI you must use
SINGle
Sequence mode.
Parameters:
<Mode>
SINGle
Each measurement is performed once (regardless of the chan-
nel's sweep mode), considering each channels' sweep count,
until all measurements in all active channels have been per-
formed.
CONTinuous
The measurements in each active channel are performed one
after the other, repeatedly (regardless of the channel's sweep
mode), in the same order, until the Sequencer is stopped.
CDEFined
First, a single sequence is performed. Then, only those channels
in continuous sweep mode (
INIT:CONT ON
) are repeated.
*RST:
CONTinuous
Example:
SYST:SEQ ON
Activates the Sequencer.
INIT:SEQ:MODE SING
Sets single sequence mode so each active measurement will be
performed once.
INIT:SEQ:IMM
Starts the sequential measurements.