Command Reference
R&S
®
ZNL/ZNLE
598
User Manual 1178.5966.02 ─ 07
Parameters:
<Boolean>
ON - time gate enabled
OFF - time gate disabled
*RST:
OFF
Example:
*RST; :CALC:TRAN:TIME:STAT?
CALC:FILT:TIME:STAT?
Reset the instrument, activating a frequency sweep, and query
whether the default trace is displayed in the time domain and
whether the time gate is enabled. The response to both queries
is
0
.
Manual operation:
See
CALCulate<Chn>:FILTer[:GATE]:TIME:WINDow
<TimeGate>
Selects the time gate to be applied to the time domain transform.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<TimeGate>
RECT | HAMMing | HANNing | BOHMan | DCHebyshev
RECT - steepest edges (rectangle)
HANN - normal gate (Hann)
HAMMing - steep edges (Hamming)
BOHMan - minimum flatness (Bohman)
DCHebyshev - arbitrary gate shape (Dolph-Chebychev)
*RST:
HANN
Example:
See
CALCulate<Chn>:FILTer[:GATE]:TIME:DCHebyshev
Manual operation:
See
CALCulate<Chn>:FORMat...
The
CALCulate:FORMat...
commands determine the post-processing of the mea-
sured data in order to obtain various display formats.
CALCulate<Chn>:FORMat
<Type>
Defines how the measured result at any sweep point is post-processed and presented
in the graphical display.
Note:
The analyzer allows arbitrary combinations of display formats and measured
quantities; see
Chapter 9.8, "Format Softtool"
CALCulate<Ch>:PARameter...
commands. Nevertheless, it is advisable to check
which display formats are generally appropriate for an analysis of a particular mea-
sured quantity; see
Chapter 8.2.3.3, "Measured Quantities and Trace Formats"
Suffix:
<Chn>
.
Channel number used to identify the active trace
VNA Command Reference
www.allice.de
Allice Messtechnik GmbH