Command Reference
R&S
®
ZNL/ZNLE
544
User Manual 1178.5966.02 ─ 07
EXTernal
The external reference from the "REF INPUT 10
MHZ" connec-
tor is used; if none is available, an error flag is displayed in the
status bar
Example:
ROSC:SOUR EXT
10.3.4.2
Calibration and Checks
The following commands control calibration and checks on the R&S
ZNL/ZNLE.
DIAGnostic:SERVice:INPut:MC:CFRequency
................................................................... 545
DIAGnostic:SERVice:INPut:MC[:DISTance]
...................................................................... 545
DIAGnostic:SERVice:INPut:PULSed:CFRequency
............................................................ 545
DIAGnostic:SERVice:INPut:PULSed:MCFRequency
..........................................................546
DIAGnostic:SERVice:INPut:PULSed:WBFRequency
......................................................... 546
DIAGnostic:SERVice:INPut:RF[:SPECtrum]
......................................................................546
DIAGnostic:SERVice:INPut[:SELect]
................................................................................ 547
DIAGnostic:SERVice:STESt:RESult?
............................................................................... 547
SOURce<si>:TEMPerature:FRONtend
.............................................................................547
CALibration[:ALL]
<UseData>
This command initiates a calibration (self-alignment) routine and queries if calibration
was successful.
During the acquisition of correction data the instrument does not accept any remote
control commands.
In order to recognize when the acquisition of correction data is completed, the MAV bit
in the status byte can be used. If the associated bit is set in the Service Request
Enable (SRE) register, the instrument generates a service request after the acquisition
of correction data has been completed.
Return values:
<CalState>
ON | OFF | 0 | 1
OFF | 0
Calibration was successful.
ON | 1
Calibration was not successful.
Example:
*CLS
Resets the status management.
*SRE 16
Enables MAV bit in the Service Request Enable register.
*CAL?
Starts the correction data recording, and then a service request
is generated.
Common Instrument Commands