Command Reference
R&S
®
ZNL/ZNLE
905
User Manual 1178.5966.02 ─ 07
SYSTem:LANGuage
<Language>
Specifies the remote language for the analyzer.
Parameters:
<Language>
Command syntax for the analyzer, string variable:
'SCPI'
–
R&S
ZNL/ZNLE-specific command set: the analyzer
supports all commands described in this documentation.
'ZVR' | 'ZVABT'
–
compatibility with network analyzers of the
R&S ZVR and R&S ZVA/B/T families.
'PNA' | 'HP8510' | 'HP8720' | 'HP8753' | 'HP8714' | 'HP8530' |
'ENA'
–
compatibility with network analyzers from other manu-
facturers.
*RST:
n/a - a reset does not affect the language setting.
The factory setting is SCPI.
Example:
SYST:LANG 'PNA'
Select a PNA-compatible command set.
SYSTem:PRESet:USER:CAL
<PresetUserCal>
Selects a calibration from the calibration pool that shall be restored during a user-
defined preset.
The corresponding cal group file (
<PresetUserCal>.cal
) must be available in the
cal pool directory
C:\Users\Public\Documents\Rohde-Schwarz\ZNL\Calibration\Data\
.
Parameters:
<PresetUserCal>
Name of the cal group (i.e. the name of the cal group file without
path and extension).
Use the empty string to restore the default behavior (no preset
user cal).
Example:
MMEMORY:STORE:CORRection 1, 'Refl1
Trans123.cal'
Adds the active user calibration to the memory pool
SYSTem:PRESet:USER:CAL 'Refl1 Trans123'
Sets the newly created cal pool member as the preset user cal.
SYSTem:PRESet:USER:CAL?
Queries for the preset user cal. Returns
'Refl1 Trans123'
SYSTem:PRESet:USER:CAL ''
Restores the default behaviour of the instrument: no preset user
cal.
Manual operation:
SYSTem:SETTings:UPDate
<Activate>
Initiates an immediate update of the channel or trace settings.
VNA Command Reference
www.allice.de
Allice Messtechnik GmbH