Command Reference
R&S
®
ZNL/ZNLE
704
User Manual 1178.5966.02 ─ 07
Parameters:
<Capacitance>
Capacitance C for ground loop deembedding.
Range:
-1mF to 1 mF.
Increment: 1 fF (1E-15 F)
*RST:
1 pF (1E-12 F)
Default unit: F
Parameters for setting and query:
<CircuitModel>
SC | SG
Possible circuit models (character data); see
Example:
*RST; :CALC:TRAN:VNET:GLO:DEEM:PAR:C? SC
Query the default capacitance for ground loop deembedding.
The response is 1E-012 (1 pF).
CALC:TRAN:VNET:GLO:DEEM:PAR:C SC, 2.2E-12
Increase the capacitance to 2.2 pF.
Manual operation:
CALCulate<Ch>:TRANsform:VNETworks:GLOop:DEEMbedding:PARameters:G
<CircuitModel>, <Conductance>
CALCulate<Ch>:TRANsform:VNETworks:GLOop:DEEMbedding:PARameters:G?
<CircuitModel>
Specifies the conductance value G in the different circuit models for ground loop
embedding.
Suffix:
<Ch>
.
Channel number.
Parameters:
<Conductance>
Conductance G for the specified circuit model.
Range:
-1kS to 1 kS.
Increment: 1 nS (1E-9 S)
*RST:
0 S
Default unit: Siemens (SI unit symbol: S)
Parameters for setting and query:
<CircuitModel>
SG
Possible circuit models (character data); see
Manual operation:
CALCulate<Ch>:TRANsform:VNETworks:GLOop:DEEMbedding:PARameters:L
<CircuitModel>, <Inductance>
CALCulate<Ch>:TRANsform:VNETworks:GLOop:DEEMbedding:PARameters:L?
<CircuitModel>
Specifies the inductance value in the different circuit models for ground loop deembed-
ding.
VNA Command Reference
www.allice.de
Allice Messtechnik GmbH