Command Reference
R&S
®
ZNL/ZNLE
695
User Manual 1178.5966.02 ─ 07
Parameters:
<CircuitModel>
FIMPort | STSL | STSC | SLST | SCST | CSSL | LSSC | CSSC |
LSSL | SLCS | SCLS | SCCS | SLLS | STSG | SGST | GSSL |
LSSG | GSSG | SLGS | SGLS | SGGS
Possible circuit models (character data); see
*RST:
CSSL
Example:
CALCulate<Ch>:TRANsform:VNETworks:BALanced:
Manual operation:
CALCulate<Ch>:TRANsform:VNETworks:BALanced:EMBedding<LogPt>[:STATe]
<Boolean>
Enables or disables the embedding function for balanced ports. It is allowed to change
the circuit model and its parameters while embedding is enabled.
Suffix:
<Ch>
.
Channel number
<LogPt>
Logical port number (balanced port)
Parameters:
<Boolean>
ON | OFF - embedding active or inactive
*RST:
OFF
Example:
*RST; SOUR:LPOR1 1,2; LPOR2 3,4
Define a balanced port configuration.
CALC:TRAN:VNET:BAL:EMB:TND CSSL
Select the Serial Cs, shunt L circuit model for embedding.
CALC:TRAN:VNET:BAL:EMB:PAR:R3 CSSL, 2.2E+3; :
CALC:TRAN:VNET:BAL:EMB ON
Increase the resistance R3 for the Serial Cs, shunt L circuit
model to 2.2 kΩ and enable embedding.
Manual operation:
CALCulate<Ch>:TRANsform:VNETworks:BALanced:EMBedding<LogPt>:
PARameters:C<Cmp>
<CiruitModel>, <Capacitance>
CALCulate<Ch>:TRANsform:VNETworks:BALanced:EMBedding<LogPt>:
PARameters:C<Cmp>?
<CiruitModel>
Specifies the capacitance value C<Cmp> in the different circuit models for balanced
port embedding.
In the query form, the <Capacitance> parameter must be omitted. The command
returns the capacitance value for the specified circuit model.
Suffix:
<Ch>
.
Channel number
<LogPt>
Logical port number (balanced port)
VNA Command Reference