Example:
Suppose that the active channel setup contains an active trace
no. 1.
CALC:MARK ON
Create marker M1 and assign it to trace no. 1.
CALC:MARK:FUNC:EXEC MAX; RES?
Move the created marker to the absolute maximum of the trace
and query the stimulus and response value of the search result.
Usage:
Setting only
Manual operation:
The analyzer provides the following search modes:
Mode
Find...
MAXimum
Absolute maximum in the search range (see
MINimum
Absolute maximum in the search range
RPEak
Next valid peak to the right of the current marker position
LPEak
Next valid peak to the left
NPEak
Next highest or lowest value among the valid peaks (next peak)
TARGet
Target value (see
CALCulate<Chn>:MARKer<Mk>:TARGet
RTARget
Next target value to the right of the current marker position
LTARget
Next target value to the left
BFILter
Bandfilter search. The results are queried using
.
MMAXimum or
MMINimum
Multiple peak search
SPRogress
Sweep progress
CALCulate<Chn>:MARKer<Mk>:FUNCtion:RESult?
Returns the result (stimulus and response value) of a search started by means of
CALCulate<Chn>:MARKer<Mk>:FUNCtion:EXECute
ted before the command is enabled.
Suffix:
<Chn>
.
Channel number used to identify the active trace
<Mk>
This numeric suffix is ignored and may be set to any value.
Example:
CALCulate<Chn>:MARKer<Mk>:FUNCtion:EXECute
Usage:
Query only