[SENSe:]CORRection:COLLect:CHANnels:ALL
<Boolean>
Enables calibration of all channels in the active channel setup. The command is valid
for the following calibration methods:
●
Manual system error correction
●
Automatic system error correction (cal unit)
Parameters:
<Boolean>
ON
–
calibrate all channels.
OFF
–
calibrate the active channel only.
Example:
See
[SENSe<Ch>:]CORRection:COLLect:SAVE:
Manual operation:
See
[SENSe:]CORRection:COLLect:CHANnels:MCTYpes
<Boolean>
Toggles the
M
ultiple
C
alibration
TYpes
mode that allows calibrate a subset of the avail-
able channels using channel-specific ports and calibration types (see
"Parallel Calibration of Multiple Channels"
Enable the MCTYpes mode before defining the calibrations to be performed (using
[SENSe<Ch>:]CORRection:COLLect:METHod:DEFine
, disable it after the calibra-
tions were saved (using
[SENSe<Ch>:]CORRection:COLLect:SAVE:SELected[:
once
). Also make sure to call
SENSe<Ch>:CORRection:COLLect:SAVE:SELected
with
<Ch>
pointing to one of
the calibrated channels.
Parameters:
<Boolean>