Command Reference
R&S
®
ZNL/ZNLE
850
User Manual 1178.5966.02 ─ 07
THRoughs
Measued Throughs: comma-separated list of directed port num-
ber pairs <pno1-pno2>
TSTamp
Timestamp in UTC
LTSTamp
Timestamp in local time
CKIT
If created with FW version 2.30 or higher, for each port the name
of the used calibration kit is stored with a calibration.
The query can return one of the following:
the cal kit name
(string)
if available and unique
'Multiple'
if cal kit names are available but not unique, i.e. if <PhyPt> was
not specified and multiple cal kits were used (multi-port cals with
different connector types, merged cals)
'Unknown'
if cal kit names are not available (cals created with a FW version
< 2.30)
<Index>
Not used
Example:
SENSe:CORRection:DATA:PARameter?
Unrestricted query. Result looks like this:
100000,8500000000,201,-10,LIN
SENSe:CORRection:DATA:PARameter? STARt
...
SENSe:CORRection:DATA:PARameter? STYPe
Query settings one by one.
SENSe:CORRection:DATA:PARameter? RATTenuation
Query receiver attenuations. Result looks like this:
1,0.000000,2,0.000000
Usage:
Query only
Manual operation:
See
[SENSe<Ch>:]CORRection:DATA:PARameter:COUNt?
Gets the number of active system error corrections for channel <Ch>.
For the R&S
ZNL/ZNLE the result is 0 (uncalibrated) or 1 (calibrated). The properties of
the active system error correction can be queried using
.
Suffix:
<Ch>
.
Channel number
Usage:
Query only
VNA Command Reference
www.allice.de
Allice Messtechnik GmbH