
Common Instrument Functions
R&S
®
ZNL/ZNLE
188
User Manual 1178.5966.02 ─ 07
5. To enable the keys of the R&S
ZNL/ZNLE again, switch the instrument to local
mode (
GTL
- Go to Local), i.e. deactivate the
REN
line of the remote control inter-
face.
If the instrument is operated exclusively in remote control, it is recommended that you
switch off the display. For details see
7.4.5.9
How to Return to Manual Operation
Before you switch back to manual operation, all remote command processing must be
completed. Otherwise, the instrument will switch back to remote control immediately.
► Select the "Local" softkey, or use the following GPIB command:
status = viGpibControlREN(vi, VI_GPIB_REN_ADDRESS_GTL)
If you select the "Local" softkey while a self-alignment or a self-test is still running
(which was started remotely), the instrument only returns to the manual operation state
when the alignment or test is completed.
Network and Remote Settings
www.allice.de
Allice Messtechnik GmbH