Command Reference
R&S
®
ZNL/ZNLE
728
User Manual 1178.5966.02 ─ 07
Parameters for setting and query:
<CircuitModel>
CSL | LSC | CSC | LSL | SLC | SCL | SCC | SLL | SHLC | GSL |
LSG | SLG | SGL
Possible circuit models (character data); see
.
Example:
*RST; :CALC:TRAN:VNET:SEND:EMB:PAR:R1? CSL; R2?
CSL
Query the default resistances for the Serial C, shunt L circuit
model. The response is 10000000; 0.
CALC:TRAN:VNET:SEND:EMB:PAR:R2 CSL, 2.2E+3
Increase the resistance R2 to 2.2 kΩ.
Manual operation:
CALCulate<Ch>:TRANsform:VNETworks:SENDed:EMBedding<PhyPt>:
TNDefinition
<CircuitModel>
Selects the circuit model for single ended port embedding.
Suffix:
<Ch>
.
Channel number
<PhyPt>
Physical port number
Parameters:
<CircuitModel>
FIMPort | CSL | LSC | CSC | LSL | SLC | SCL | SCC | SLL |
SHLC | GSL | LSG | GSG | SLG | SGL | SGG
Possible circuit models (character data); see
*RST:
CSL
Example:
CALCulate<Ch>:TRANsform:VNETworks:SENDed:
Manual operation:
CALCulate<Ch>:TRANsform:VNETworks:SENDed:EMBedding<PhyPt>:
PARameters:DATA
<Interchange>, <arg1>
Defines a dembedding network for a single-ended port based on the given S-Parame-
ter traces.
Use
●
CALCulate<Ch>:TRANsform:VNETworks:SENDed:EMBedding<PhyPt>:
TNDefinition
to select the adequate circuit model
before
executing this com-
mand.
●
MMEMory:LOAD:VNETworks<Ch>:SENDed:EMBedding<PhyPt>
to load circuit
data from a Touchstone file located at the R&S
ZNL/ZNLE's file system.
Suffix:
<Ch>
.
Channel number
VNA Command Reference
www.allice.de
Allice Messtechnik GmbH