Getting Started
R&S
®
ZNL/ZNLE
60
User Manual 1178.5966.02 ─ 07
6. The "Calibration" dock widget indicates the standard measurements that make up
a "Trans Norm" calibration.
Select "Through (mm)" to initiate the measurement of the connected Through stan-
dard. Measuring the isolation between ports 1 and 2 is optional. Skip it for now.
The analyzer performs a calibration sweep for the measured quantity S
21
. The
magnitude and phase of the result is displayed in two diagrams, together with the
expected typical result for a Through standard. The similarity of real and expected
traces indicates that the Through standard has been properly connected. After the
R&S
ZNL/ZNLE has completed the calibration sweep and calculated the correction
data, the "Apply" button is enabled.
Trying Out the Instrument