VNA GUI Reference
R&S
®
ZNL/ZNLE
490
User Manual 1178.5966.02 ─ 07
Figure 9-20: Measure Fixture dialog: Direct Compensation (and Prompt for Each Port)
"Direct Compensation" data files are standard trace files, containing reflection parame-
ter traces for the related port and standard:
Table 9-3: Direct Compensation data
Prompt for Each Port
File Type
Description
disabled
csv
One csv trace file per standard,
containing reflection traces for all
selected ports; see
ter 8.4.2.2, "ASCII (*.csv) Files"
enabled
s1p
One 1-port Touchstone file per
standard and port (see
ter 8.4.2.1, "Touchstone Files"
Tip:
Remote control provides additional flexibility. You can:
●
Measure the same ports repeatedly without changing the standards and attribute
the results to different channels.
●
Calculate the compensation data for different ports, using mixed Open and Short
standards.
Refer to the program example for
[SENSe<Ch>:]CORRection:COLLect:
.
Remote command:
[SENSe<Ch>:]CORRection:COLLect:FIXTure:STARt
[SENSe<Ch>:]CORRection:COLLect:FIXTure[:ACQuire]
[SENSe<Ch>:]CORRection:COLLect:FIXTure:SAVE
[SENSe<Ch>:]CORRection:COLLect:FIXTure:EXPort
[SENSe<Ch>:]CORRection:COLLect:FIXTure:IMPort
9.14.3
One Way Loss Tab
Defines loss parameters for each physical port.
Use the complementary dock widget to activate or deactivate length/loss compensation
for selected ports (see
Chapter 9.14.1.2, "Offset Panel"
Offset Embed Softtool