VNA GUI Reference
R&S
®
ZNL/ZNLE
447
User Manual 1178.5966.02 ─ 07
2.
"Connectors and Cal Kits":
For all ports you want to calibrate, select the connec-
tor type, gender and cal kit. If necessary, import a calibration kit.
3.
"Calibration":
Acquire measurement data for the required ports or port pairs and
the required standards. Finally, decide whether to apply the resulting system error
correction or to discard it.
Background and related information
●
on page 242 for background information
●
If the active channel is already equipped with a system error correction, the "Cali-
bration Setting" wizard loads the underlying setup. If the calibration setup is not
changed and sweep data are available from previous calibrations, the existing sys-
tem error correction can be optimized without repeating the measurement of all
standards.
●
When you apply the new system error correction, the current calibration is replaced
and discarded.
To persist the current calibration, you can transfer it to the calibration "Pool" using
the
●
The active system error correction data can be read (and modified) using the
remote control command
.
Step 1: Ports and Type
Allows you to select the test ports to be calibrated and the calibration type to be used.
Background information
Refer to
Chapter 8.5.1, "Calibration Types"
Ports
Selects the test ports to be calibrated.
Cal Softtool