![Rohde & Schwarz 1323.0012.03 User Manual Download Page 319](http://html1.mh-extra.com/html/rohde-and-schwarz/1323-0012-03/1323-0012-03_user-manual_1477777319.webp)
VNA GUI Reference
R&S
®
ZNL/ZNLE
319
User Manual 1178.5966.02 ─ 07
The table in the "Trigger Manager" dialog contains several editable (white) or non-edit-
able (gray) columns. All settings are analogous to the general trigger settings in the
. Refer to the following sections:
●
●
"FreeRun / External / Manual / Multiple Triggers"
●
●
The column "Slope / Level 2" is only available, if the instrument is equipped with a sec-
ond external trigger input. This second external trigger input is provided by the Aux.
Port connector, which is part of option R&S
FPL1-B5 "Additional Interfaces" (R&S
ZNL
only).
... /Source/ ...
The table defines all settings related to "Multiple Triggers" (see
on page 315). For all measurement sequences the following trigger sources are
available:
●
"Free Run" means that a trigger signal is not required
●
"External 1" is the external trigger signal fed in at either the Trigger In connector on
the rear panel or pin 2 of the Aux. Port connector.
●
"External 2" is the external trigger signal fed in at pin 25 of the Aux. Port connector
on the rear panel.
●
"External 1 and 2"/"External 1 or 2" means that the measurement sequence is initi-
ated after the analyzer has received an event from external trigger signal 1 and/or
2. A
must be specified for both trigger signals.
●
Manual
means that the trigger event is generated manually, by selecting the
For example, a new sweep can be triggered by an external trigger no. 1, while the indi-
vidual sweep points are triggered by external trigger no. 2. External trigger 1 is ignored
if a sweep is running, external trigger 2 is ignored if there is no running sweep. In the
figure below a sweep comprises 5 measurement points and dotted arrows depict
ignored trigger events.
Remote command:
TRIGger<Ch>[:SEQuence]:MULTiple:SOURce
TRIGger<Ch>[:SEQuence]:MULTiple:SLOPe<Num>
TRIGger<Ch>[:SEQuence]:MULTiple:HOLDoff
Sweep Softtool
www.allice.de
Allice Messtechnik GmbH