[SENSe<Ch>:]FREQuency:SEGMent:AXIS
...................................................................... 861
....................................................................................861
................................................................................... 862
....................................................................................862
[SENSe<Ch>:]FREQuency[:CW]
<FixedFreq>
[SENSe<Ch>:]FREQuency:FIXed
<FixedFreq>
Defines the fixed (Continuous Wave, CW) frequency for all sweep types operating at
fixed frequency ("Time", "CW Mode"). The two command forms
[SENSe<Ch>:]FREQuency[:CW]
and
[SENSe<Ch>:]FREQuency:FIXed
are
equivalent.
The frequency ranges are listed in
.
Note:
[SENSe<Ch>:]FREQuency[:CW]|:FIXed
is equivalent to
SOURce<Ch>:FREQuency<PhyPt>[:CW]|:FIXed
. Source and receiver frequency
are always equal; the four commands overwrite each other.
Suffix:
<Ch>
.
Channel number
Parameters:
<FixedFreq>
Fixed stimulus and analyzer frequency
*RST:
1 GHz
Default unit: Hz
Example:
FUNC "XTIMe:POW:A1"
Activate a time sweep and select the wave quantity a
1
as mea-
sured parameter for channel and trace no. 1.
FREQ:CW 100MHz
Set the CW frequency to 100 MHz.
Manual operation:
See
[SENSe<Ch>:]FREQuency:CENTer
<CenterFreq>
Defines the center of the measurement and display range for a frequency sweep
(sweep range). The default center frequency is the center of the analyzer's maximum
frequency range: (f
MIN
+ f
MAX
)/2. The range is listed in table
.
Suffix:
<Ch>
.
Channel number
Parameters:
<CenterFreq>
Center frequency of the sweep
Increment: 0.1 kHz
Default unit: Hz