Getting Started
R&S
®
ZNL/ZNLE
57
User Manual 1178.5966.02 ─ 07
output port. The trace settings allow you to select the measured quantities and display
formats, depending on what you want to learn from the data. A minimum of two ana-
lyzer test ports are required for transmission measurements.
In the following example, the analyzer is set up for a two-port transmission measure-
ment. A frequency sweep range is selected, the instrument is calibrated and the mea-
surement result is analyzed using various display formats.
Connecting the Instrument for Transmission Measurements
To prepare a transmission measurement, you have to connect your DUT (which for
simplicity we assume to have appropriate connectors) in-between a pair of analyzer
test ports. It is recommended that you preset the R&S
ZNL/ZNLE to start from a well-
defined instrument state.
1. Connect the DUT between test ports 1 and 2 of the network analyzer as shown
above.
2. Switch on the instrument and start the VNA application.
Proceed as described in
Chapter 5.1.1.8, "Switching the Instrument On and Off"
3. Use the [Preset] key to restore a well-defined instrument state.
The analyzer is now set to its default state. The default measured quantity is the
transmission S-parameter S
21
.
Trying Out the Instrument