Command Reference
R&S
®
ZNL/ZNLE
727
User Manual 1178.5966.02 ─ 07
<PhyPt>
Physical port number
<Cmp>
Number of inductance in circuit model. The total number of
inductances depends on the selected circuit model.
Parameters:
<Inductance>
Inductance L<Cmp> for the specified circuit model.
Range:
-1H to 1 H.
Increment: 1 pH (1E-12 H)
*RST:
1 nH (1E-9 H)
Default unit: H
Parameters for setting and query:
<CircuitModel>
CSL | LSC | LSL | SLC | SCL | SLL | SHLC | GSL | LSG | SLG |
SGL
Possible circuit models (character data); see
Example:
*RST; :CALC:TRAN:VNET:SEND:EMB:PAR:L1? SLL
Query the default inductance L1 for the Shunt L, serial L circuit
model. The response is 1E-009 (1 nH).
CALC:TRAN:VNET:SEND:EMB:PAR:L1 SLL, 2.2E-9
Increase the inductance to 2.2 nH.
Manual operation:
See
CALCulate<Ch>:TRANsform:VNETworks:SENDed:EMBedding<PhyPt>:
PARameters:R<Cmp>
<CircuitModel>, <Resistance>
CALCulate<Ch>:TRANsform:VNETworks:SENDed:EMBedding<PhyPt>:
PARameters:R<Cmp>?
<CircuitModel>
Specifies the resistance value R<Cmp> in the different circuit models for single ended
port embedding.
In the query form, the <Resistance> parameter must be omitted. The command returns
the resistance value for the specified circuit model.
Suffix:
<Ch>
.
Channel number
<PhyPt>
Physical port number
<Cmp>
Number of resistance in circuit model. The total number of resis-
tances depends on the selected circuit model.
Parameters:
<Resistance>
Resistance R<Cmp> for the specified circuit model.
Range:
-10 MΩ to 10 MΩ.
Increment: 1 mΩ (1E-3 Ω)
*RST:
0 Ω for all resistances connected in series with an
inductance. 10 MΩ for all resistances connected in
parallel with a capacitance
Default unit: Ohm
VNA Command Reference