VNA GUI Reference
R&S
®
ZNL/ZNLE
342
User Manual 1178.5966.02 ─ 07
9.6.7
Distance to Fault Tab
With option K3, the R&S
ZNL/ZNLE can locate faults and discontinuities on cables and
transmission lines. The measured S-parameter trace is mathematically converted to
the time domain, and represented in the distance domain. On the resulting trace, faults
appear as peaks above a certain level.
The default measurement logic implemented on the "Distance to Fault" (DTF) tab focu-
ses on the canonical case where the reflections of cable connected to Port 2 shall be
observed. The x-axis of the DTF trace (derived from S22) shows the distance from the
calibrated reference plane.
●
If option K3 is not installed on the R&S
ZNL/ZNLE, the "Distance to Fault" tab is
hidden.
●
In order to achieve the best distance resolution, set the sweep span to its maxi-
mum possible value and use the
sweep points.
Background information
Chapter 8.7.2, "Distance to Fault Measurements"
9.6.7.1
Controls on the Distance to Fault Tab
The "Distance to Fault" tab provides controls that allow to enable and configure a stan-
dard Distance to Fault (DtF) measurement.
Trace Softtool