Command Reference
R&S
®
ZNL/ZNLE
697
User Manual 1178.5966.02 ─ 07
<Port>
Port assignment for two 2-port networks:
1
- Port 1
2
- Port 2
This parameter is ignored for 4-port networks.
Setting parameters:
<Interchange>
FPORts | IPORts | SGATes | SINCreasing
FPORts (or omitted)
Standard port sequence (odd port numbers towards VNA, even
port numbers towards DUT)
IPORts
– Two-port networks: inverted port sequence (network port 2
towards VNA, network port 1 towards DUT)
– Four-port networks: increasing port sequence (low port num-
bers towards VNA, high port numbers towards DUT)
SGATes
Swapped gates (even port numbers towards VNA, odd port
numbers towards DUT)
SINCreasing
Swapped increasing port sequence (high port numbers towards
VNA, low port numbers towards DUT)
<arg1>
<block_data>
Content of a Touchstone file (
*.s2p
or
*.s4p
) in IEEE488.2
Usage:
Setting only
CALCulate<Ch>:TRANsform:VNETworks:BALanced:EMBedding<LogPt>:
PARameters:G<Cmp>
<CircuitModel>, <Conductance>
CALCulate<Ch>:TRANsform:VNETworks:BALanced:EMBedding<LogPt>:
PARameters:G<Cmp>?
<CircuitModel>
Specifies the conductance value G<Cmp> in the different circuit models for balanced
port embedding.
In the query form, the <Conductance> parameter must be omitted. The command
returns the capacitance value for the specified circuit model.
Suffix:
<Ch>
.
Channel number
<LogPt>
Logical port number (balanced port)
<Cmp>
Number of conductance in circuit model. The total number of
conductances depends on the selected circuit model.
VNA Command Reference
www.allice.de
Allice Messtechnik GmbH