Example:
CALC4:PAR:SDEF 'Ch4Tr1', 'S11'
Create channel 4 (channel suffix 4) and a trace named "Ch4Tr1" to measure the input
reflection coefficient S11. The trace is created but not displayed.
DISP:WIND2:STAT ON
Create diagram area no. 2 (window suffix 2).
DISP:WIND2:TRAC9:FEED 'CH4TR1
'
Display the generated trace (identified by its name "Ch4Tr1") in diagram area no. 2
(window suffix 2), assigning the trace number 9 (trace suffix 9) to it.
10.4.2.2
Active Traces in Remote Control
In manual control there is always exactly one active trace, irrespective of the number of
channels and traces defined. The "active channel" contains the active trace; see
In remote control, each channel contains an active trace (unless the channel contains
no trace at all). This principle actually simplifies the remote control command syntax,
because it allows the active trace in a particular channel to be referenced by means of
the channel suffix. No additional trace identifier is needed; there is no need either to
distinguish channel and trace settings using mnemonics or suffixes.
The active traces are handled as follows:
●
After a preset (
*RST
), the analyzer displays a single diagram area with the default
trace no. 1 named
Trc1
. The trace is active in manual and in remote control.
●
In manual control, a new, added trace automatically becomes the active trace. To
select another trace as the active trace, tap inside the trace list.
●
In remote control, a new trace added via
CALCulate<Ch>:PARameter:SDEFine
'<trace_name>', '<parameter>'
also becomes the active trace. To select
another trace as the active trace, use (
CALCulate<Ch>:PARameter:SELect
'<trace_name>'
).
●
The active traces for manual and remote control may be different.
Example:
*RST
Reset the analyzer, creating channel no. 1 with the default trace "Trc1". The trace is
displayed in diagram area no. 1.
CALC1:PAR:SDEF 'Trc2', 'S11'
;
DISP:WIND:TRAC2:FEED 'Trc2'
Create a new trace named "Trc2", assigned to channel no. 1 (the suffix 1 after
CALC
,
may be omitted), and display the trace. The new trace automatically becomes the
active trace for manual and for remote control. To check this, tap "Trace – Marker –
Marker 1" to create a marker. The marker is assigned to "Trc2". Delete all markers
("Trace – Marker – All Markers Off").
CALC1:MARK ON