![Rohde & Schwarz 1323.0012.03 User Manual Download Page 673](http://html1.mh-extra.com/html/rohde-and-schwarz/1323-0012-03/1323-0012-03_user-manual_1477777673.webp)
Command Reference
R&S
®
ZNL/ZNLE
673
User Manual 1178.5966.02 ─ 07
Usage:
Event
Manual operation:
See
CALCulate<Chn>:TRANsform:TIME:LPFRequency
Calculates the harmonic grid for low pass time domain transforms, keeping the stop
frequency and the number of points.
Tip:
CALCulate<Chn>:TRANsform:TIME:LPASs
other algorithms for calculating the grid.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Example:
See
CALCulate<Chn>:TRANsform:TIME[:TYPE]
Usage:
Event
Manual operation:
See
CALCulate<Chn>:TRANsform:TIME:RESolution:EFACtor
<REfactor>
Defines the resolution enhancement factor for the time domain transform.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<REfactor>
Resolution enhancement factor.
Range:
1 to 10.
Increment: 0.1
*RST:
1 (no resolution enhancement)
Example:
See
CALCulate<Chn>:TRANsform:TIME:LPASs:DCSParam
Manual operation:
See
CALCulate<Chn>:TRANsform:TIME:SPAN
<Span>
Defines the time span of the diagram in time domain.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Span>
Time span of the diagram in time domain.
Range:
2E-012 s to 200 s.
Increment: 0.1 ns
*RST:
5E-009 s
Default unit: s
Example:
See
CALCulate<Chn>:TRANsform:TIME:CENTer
VNA Command Reference
www.allice.de
Allice Messtechnik GmbH