Common Instrument Functions
R&S
®
ZNL/ZNLE
175
User Manual 1178.5966.02 ─ 07
ing to the instrument, the LXI status LED blinks. The "Device Indicator" setting is
not password-protected.
The most important control elements in the navigation pane of the browser interface
are the following:
●
"LAN Configuration" opens the menu with configuration pages.
●
"Status" displays information about the LXI status of the instrument.
●
"Help > Glossary" opens a document with a glossary of terms related to the LXI
standard.
LAN Configuration
The LAN configuration consists of three parts:
●
"IP configuration" provides all mandatory LAN parameters.
●
"Advanced LAN Configuration" provides LAN settings that are not declared manda-
tory by the LXI standard.
●
"Ping Client" provides the ping utility to verify the connection between the instru-
ment and other devices.
IP Configuration
The "LAN Configuration > IP configuration" web page displays all mandatory LAN
parameters and allows their modification.
The "TCP/IP Mode" configuration field controls how the IP address for the instrument
gets assigned (see also
"How to Assign the IP Address"
For the manual configuration mode, the static IP address, subnet mask, and default
gateway are used to configure the LAN. The automatic configuration mode uses DHCP
server or Dynamic Link Local Addressing (Automatic IP) to obtain the instrument IP
address.
Changing the LAN configuration is password-protected. The default password is
Lxi-
WebIfc
(notice upper and lower case characters).
You can change the LXI password in the "N Remote" dialog box, see .
Advanced LAN Configuration
The "LAN Configuration > Advanced LAN Configuration" parameters are used as fol-
lows:
●
The "Negotiation" configuration field provides different Ethernet speed and duplex
mode settings. In general, the "Auto Detect" mode is sufficient.
●
"ICMP Ping" must be enabled to use the ping utility.
●
"VXI-11" is the protocol that is used to detect the instrument in the LAN. According
to the standard, LXI devices must use VXI-11 to provide a detection mechanism;
other additional detection mechanisms are permitted.