VNA GUI Reference
R&S
®
ZNL/ZNLE
431
User Manual 1178.5966.02 ─ 07
Search Config
.
Remote command:
CALCulate<Chn>:MARKer<Mk>:FUNCtion:EXECute
MMAXimum | MMINimum
Threshold Settings
Defines a threshold for multiple peak searches and activates it.
If "Active", only maxima above and minima below the configured "Threshold" are con-
sidered.
The unit is derived from the active trace format and cannot be modified. Changing the
trace format resets the "Threshold" to a format-specific default value.
Remote command:
CALCulate<Chn>:MARKer:MPEak:THReshold
CALCulate<Chn>:MARKer:MPEak:THReshold:STATe
Excursion Settings
Defines a minimum excursion for multiple peak searches and activates it.
If "Active", only peaks with an excursion above the configured "Excursion" value are
considered. By definition, the excursion of a peak is the smaller of the absolute differ-
ences in measured values from the adjoining peaks of opposite polarity.
The unit is derived from the active trace format and cannot be modified. Changing the
trace format resets the "Excursion" to a format-specific default value.
Remote command:
CALCulate<Chn>:MARKer:MPEak:EXCursion
CALCulate<Chn>:MARKer:MPEak:EXCursion:STATe
9.11.3
Target Search Tab
The "Target Search" functions use markers to locate trace points with a specific
response value ("Target Value"). The functions are unavailable if the active trace con-
tains no markers (e.g. after "All Markers Off").
Marker Search Softtool
www.allice.de
Allice Messtechnik GmbH