Command Reference
R&S
®
ZNL/ZNLE
514
User Manual 1178.5966.02 ─ 07
<ChannelName2>
String containing the new channel setup name.
Note that you cannot assign an existing channel setup name to
a new channel setup; this will cause an error.
Channel names can have a maximum of 31 characters, and
must be compatible with the Windows conventions for file
names. In particular, they must not contain special characters
such as ":", "*", "?".
Usage:
Setting only
INSTrument[:SELect]
<ChannelType> | <ChannelName>
This command activates a new channel setup with the defined channel setup type, or
selects an existing channel setup with the specified name.
Also see
●
Parameters:
<ChannelType>
Channel type of the new channel setup.
For a list of available channel setup types see
<ChannelName>
String containing the name of the channel setup.
Example:
INST IQ
INST 'MyIQSpectrum'
Selects the channel setup named 'MyIQSpectrum' (for example
before executing further commands for that channel setup).
Manual operation:
See
10.3.2.2
Performing a Sequence of Measurements
The following commands control the sequencer.
For details on the Sequencer see
Chapter 6.4, "Running a Sequence of Measure-
............................................................................................ 514
....................................................................................... 515
.............................................................................................515
INITiate:SEQuencer:ABORt
This command stops the currently active sequence of measurements. The Sequencer
itself is not deactivated, so you can start a new sequence immediately using
To deactivate the Sequencer use
Usage:
Setting only