Messages in the dialog
An information message (or error message) is displayed if one of the following hap-
pens:
●
One of the selected calibration kits is described by ideal kit parameters or typical
values.
●
One of the selected calibration kits does not contain all standards that are required
for the previously selected calibration type.
●
Different connector types are defined at the ports but the selected calibration type
requires uniform connectors.
●
A cal kit standard does not cover the entire calibrated frequency range.
Figure 9-13: Calibration Setting Wizard, Step 2: Connectors and Cal Kits
The upper part of the panel shows the ports and the calibration type selected in
. The lower part gives access to the connector and cal kit settings.
Connector / Gender
Defines the connector types and genders of the ports to be calibrated. For symmetric
(sexless) connector types (e.g. 7 mm / PC7), "Gender" is unavailable.
If "Same Connector All Ports" is active, the connector types at all ports (but not their
gender) are always adjusted to the current selection. If "Same Gender All Ports" is
active, the genders at all ports are always adjusted to the current selection.
User-defined connectors can be added or removed in the
Remote command:
[SENSe<Ch>:]CORRection:COLLect:CONNection<PhyPt>
[SENSe<Ch>:]CORRection:COLLect:SCONnection<PhyPt>
[SENSe<Ch>:]CORRection:CONNection
[SENSe<Ch>:]CORRection:CONNection:CATalog?
[SENSe<Ch>:]CORRection:CONNection:DELete