![Rohde & Schwarz 1323.0012.03 User Manual Download Page 435](http://html1.mh-extra.com/html/rohde-and-schwarz/1323-0012-03/1323-0012-03_user-manual_1477777435.webp)
VNA GUI Reference
R&S
®
ZNL/ZNLE
435
User Manual 1178.5966.02 ─ 07
Remote command:
CALCulate<Chn>:MARKer<Mk>:BWIDth
Bandpass Ref to Max
Activates the search for a bandpass region on the active trace and activates
The located bandpass region is the tallest peak in the search range with a minimum
excursion as specified by the "Bandwidth" parameter.
If "Bandpass Ref to Max" is selected, the analyzer uses (or creates) the four markers
"M1" to "M4" to locate the
bandpass region
.
●
"M1" indicates the maximum of the peak ("Max").
●
"M2" indicates the point on the left edge of the peak where the trace value is equal
to the maximum minus the bandwidth factor ("Lower Edge").
●
"M3" indicates the point on the right edge of the peak where the trace value is
equal to the maximum minus the bandwidth factor ("Upper Edge").
●
"M4" indicates the center of the peak. Depending on a system setting, the center is
either calculated as the geometric or the arithmetic mean of the "Lower Edge" and
"Upper Edge" frequencies (see
"Geometric Calculation of Bandfilter Center "
The search results are displayed in the movable "Bandfilter" info field.
Remote command:
CALCulate<Chn>:MARKer<Mk>:FUNCtion:BWIDth:MODE BPASs
CALCulate<Chn>:MARKer<Mk>:FUNCtion:EXECute BFILter
CALCulate<Chn>:MARKer:SEARch:BFILter:RESult[:STATe]:AREA
Bandpass Ref to Mkr
Activates the search for a bandpass region on the active trace and activates
starting at the position of the active marker. A bandpass region is the closest peak in
the evaluation range that has a minimum excursion as specified by the "Bandwidth"
parameter.
In contrast to a "Bandpass Ref to Max", the "Bandpass Ref to Mkr" does not change
the position of the active markers. The search results are displayed in the movable
"Bandfilter" info field.
Marker Search Softtool
www.allice.de
Allice Messtechnik GmbH