Tip:
Drag and drop the network symbols in horizontal or vertical direction to switch to
the next symbol.
Remote command:
CALCulate<Ch>:TRANsform:VNETworks:DIFFerential:EMBedding<LogPt>:
TNDefinition
CALCulate<Ch>:TRANsform:VNETworks:DIFFerential:EMBedding<LogPt>:
PARameters:L<Cmp>
CALCulate<Ch>:TRANsform:VNETworks:DIFFerential:EMBedding<LogPt>:
PARameters:R<Cmp>
CALCulate<Ch>:TRANsform:VNETworks:DIFFerential:EMBedding<LogPt>:
PARameters:C<Cmp>
CALCulate<Ch>:TRANsform:VNETworks:DIFFerential:EMBedding<LogPt>:
PARameters:G<Cmp>
Select File...
"Select File..." is enabled as long as the "2-Port Data" network is selected. This net-
work is defined by its S-parameters stored in a two-port Touchstone file (
*.s2p
). No
additional parameters are required.
Note:
The loaded file is stored in the active channel setup. Persisted channel setups
contain the full (de-)embedding data so that they can be transferred to other instru-
ments.
Remote command:
MMEMory:LOAD:VNETworks<Ch>:DIFFerential:EMBedding<LogPt>
9.14.9
Config Tab
The "Config" tab provides means to configure the overall "Offset Embed" calculation.
Offset > Calculate after De-/Embed.
Changes the position of the "Offset" calculation in the "Offset Embed" calculation
chain.
If unchecked (default), the offset is calculated before de-/embedding. If checked, it is
calculated after de-/embedding.
Offset Embed Softtool
www.allice.de