Common Instrument Functions
R&S
®
ZNL/ZNLE
103
User Manual 1178.5966.02 ─ 07
.......................................................................................103
Configurable Storage and Recall
.......................................................................... 105
How to Save and Load Instrument Settings
..........................................................109
7.2.2.1
Quick Save/Quick Recall
The Quick Save and Quick Recall functions allow you to store instrument settings or
channel setups very easily and quickly in one step. Up to ten different sets of settings
can be stored to or recalled from "save sets". Each save set is identified by its storage
date and type (instrument or specific "Channel") in the display. The save sets are
stored in the
C:\Users\Public\Documents\Rohde-Schwarz\ZNL\user\QuickSave
direc-
tory, in files named
QuickSave1.dfl
to
QuickSave10.dfl
. The storage filenames
and locations cannot be changed. Only the current measurement settings are stored,
not any additional data such as traces, limit line or transducer files (see
During recall, save sets of type "Instrument" replace the settings of the entire instru-
ment. All other save sets start a new channel setup with the stored settings.
If a channel setup with the same name as the "Channel" to be restored is already
active, the name for the new channel setup is extended by a consecutive number:
Quick Save / Quick Recall Settings
Access
: "Save"/ "Open" icon in the toolbar > "Quick Save" / "Quick Recall"
Both dialog boxes are very similar and closely related.
Data Management
www.allice.de
Allice Messtechnik GmbH