Command Reference
R&S
®
ZNL/ZNLE
809
User Manual 1178.5966.02 ─ 07
[SENSe:]CORRection:CKIT:SELect
<ConnectorType>[, <CalKitName>]
Selects the calibration kit to be used, specifying its connector type and name
(optional).
Tip:
The command is suitable for connector types with arbitrary, user-defined names.
For standard connector types you can use the command
Parameters:
<ConnectorType>
Connector type, e.g. a user-defined connector type (string vari-
able).
<CalKitName>
String parameter containing the name of a calibration kit availa-
ble on the analyzer.
Example:
MMEM:LOAD:CKIT 'C:\Users\Public
\Documents\Rohde-Schwarz
\Vna\Calibration\Kits\New Kit 1.calkit'
Load the previously created cal kit file
New Kit 1.calkit
from the default cal kit directory.
CORR:CKIT:SEL 'N 50 Ohm', 'New Kit 1'
Assign the imported kit to the N 50 Ω connector type, assuming
that the cal kit name stored in
New Kit 1.calkit
reads
New
Kit 1
.
Manual operation:
See
[SENSe:]CORRection:CKIT:STANdard:CATalog?
<CalKitName>
Returns a list of all standards in a given calibration kit.
Query parameters:
<CalKitName>
Example:
See
[SENSe<Ch>:]CORRection:CONNection:CATalog?
Usage:
Query only
[SENSe:]CORRection:CKIT:STANdard:DATA?
<ConnectorType>, <CalKitName>,
<KitLabel>, <StandardType>, <DelayMode>[, <Port1>[, <Port2>]]
Returns the data of the related calibration standard (identified by <ConnectorType>,
<CalKitName>, <KitLabel> and <StandardType>) and - optionally - restricted to the
given port(s).
The delay parameter is returned according to the selected <DelayMode>; see
.
VNA Command Reference