Command Reference
R&S
®
ZNL/ZNLE
872
User Manual 1178.5966.02 ─ 07
<Selectivity>
NORMal | MEDium | HIGH
Selectivity of the IF filter. See
.
<FreqSweepMode>
STEPped
Frequency sweep mode; must be set to
Stepped
.
Example:
SEGM:ADD
Create a new sweep segment no. 1 in channel no. 1 using
default settings.
SEGM:DEF?
Query the channel settings for the new segment. A possible
response is
1000000000,2000000000,51,-300,0.0044625,0,
10000,AUTO,NORM,STEP
.
Manual operation:
[SENSe<Ch>:]SEGMent<Seg>:DEFine:SELect
<TimeRef>
Defines whether the sweep time of a new segment, i.e. numeric parameter no. 5 of the
command
[SENSe<Ch>:]SEGMent<Seg>:DEFine
, is entered as a segment sweep
time ("Segment Time") or as a measurement delay ("Meas Delay").
Suffix:
<Ch>
.
Channel number
<Seg>
Sweep segment number
Parameters:
<TimeRef>
SWTime | DWELl
SWTime - use segment sweep time.
DWELl - use measurement delay.
Example:
SEGM1:DEF:SEL DWEL
Select the measurement delay to determine the sweep time in a
new sweep segment no. 1.
SEGM1:DEF 1MHZ, 1.5MHZ, 111, -21DBM, 0.01S, 0,
10KHZ
Create a sweep segment with a sweep range between 1.0 MHz
and 1.5 MHz and a measurement delay of 10 ms.
SEGM1:SWE:TIME?
Query the sweep time in the new segment.
Manual operation:
[SENSe<Ch>:]SEGMent<Seg>:DELete:ALL
Deletes all sweep segments in the channel.
deletes a single segment.
VNA Command Reference