
Command Reference
R&S
®
ZNL/ZNLE
650
User Manual 1178.5966.02 ─ 07
Parameters:
<TraceName>
Trace name, string variable, e.g. 'Trc4'. See "Rules for trace
names" in
on page 324. Trace names must be
unique across all channels and diagrams.
<Result>
Measurement parameter (string variable); see
.
A query of a wave quantity
'xy'
returns
'xyD<n><Detector>'
, where
<n>
numbers the source
(drive) port, and
<Detector>
denotes the detector setting. A
query of a ratio
'x/y'
returns
'xD<n>/yD<m><Detector>'
,
where
<n>
and
<m>
number the source ports
Example:
CALC4:PAR:SDEF 'Ch4Tr1', 'S11'
Create channel 4 and a trace named
Ch4Tr1
to measure the
input reflection coefficient S
11
.
CALC4:PAR:MEAS 'Ch4Tr1', 'a1'
Change the measurement parameter of the trace and measure
the wave quantity a
1
.
CALC4:PAR:MEAS? 'Ch4Tr1'
Query the measured quantity. The response is
'A1D1SAM'
.
Manual operation:
CALCulate<Ch>:PARameter:MEASure:SENDed
<TraceName>, <Result>
Assigns a measurement result to an
existing
trace. The query returns the result
assigned to the specified trace (no second parameter; see example).
CALCulate<Ch>:PARameter:MEASure
, but in presence of balanced ports
this command distinguishes between
●
single-ended S-parameters for
logical
ports (
SSS<Li><Lj>
)
●
"raw" single-ended S-parameters referring to
physical
ports (
S<Pi><Pj>
)
Suffix:
<Ch>
.
Channel number
Setting parameters:
<TraceName>
Trace name
<Result>
Measured quantity
Manual operation:
CALCulate<Ch>:PARameter:PTIP
<TraceName>, <ProbeTip>
CALCulate<Ch>:PARameter:PTIP?
<TraceName>
Gets/sets the probe tip impedance for the calculation of trace <TraceName>.
Suffix:
<Ch>
.
Channel number
VNA Command Reference
www.allice.de
Allice Messtechnik GmbH